SynchroSonic US/752
The panel design of the SynchroSonic
US/752 provides simple operation of ultrasound and high volt muscle
stimulation separately or in combination through the ultrasound
transducer. The practitioner can precisely manage treatment with a
selectable frequency range, selectable independent on/off surge duration
and selectable pulse width.
The US/752 produces pulsation, tetanize,
surge and reciprocal output. The additional benefit of bipolar
application eliminates the need for a single large dispersive
electrode. Designed to promote patient assurance, the treatment
stop switch allows the patient to discontinue stimulator output at the
press of a button.
Soundhead Features and Options
All Amrex ultrasound transducers are
lightweight, ergonomically designed and sealed for underwater therapy.
Amrex offers two soundhead faceplate
sizes: the standard 2.25" diameter faceplate or the optional,
small 1.5" diameter faceplate. The QuickConnect™ Transducer
Cable System is optional with the standard soundhead. It is
provided with the small soundhead or the dual standard/small soundhead
on all Amrex ultrasound equipment.
|
Ultrasound
Features
- Lightweight standard
transducer
- Optional QuickConnect™
transducer
- Optional small transducer
with QuickConnect cable
- Cable fault alarm system
- Transducer protection
switch
- Adjustable interrupted
output mode
- 0 to 20 watt output
• 30 minute timer
Stimulator Features
- Dual channel, 4 pad
stimulator
- Dual intensity controls
- Polarity control
- Bipolar application (no
dispersive pad)
- Selectable frequency, pulse
width, on/off surge duration
- Surge and reciprocal mode
ramped
- Intensity reset circuit
- Patient treatment stop
switch
- Combination therapy
capability
- Two Year Warranty -
Generator - Parts and Labor
- One Year Warranty -
Transducer - Parts and Labor
|
|
Input Power Requirements
| Line
Voltage |
120 Vac, 60 Hz
(Special voltages
available on request)
|
Ultrasonic Generator
(Complies with FDA Regulation 21
CFR 1050.10 and FCC Regulations)
| Waveform |
amplitude modulated at
120 Hz ± 1%
|
| Output Mode |
adjustable interrupted
& non-interrupted
|
| Output
Power |
0 to 20 W total ± 20%
0 to 5 W/cm˛ ± 20%
|
| Temporal Peak/Temporal
Average Intensity |
2.8:1 ± 20%
|
| Treatment Timer |
adjustable 0 to 30
minutes
|
Applicator (Transducer)
| Crystal Material |
barium-titanate
|
| Applicator Beam Type |
collimating
|
| ERA (Effective
Radiating Area) |
4 cm˛ ± 20%
|
| BNR (Beam Nonuniformity
Ratio) |
5.5:1 maximum
|
| Small Faceplate Size |
1.5" diameter
|
| Standard Faceplate Size |
2.25" diameter
|
Caution: Federal law (USA)
restricts this device to sale by or on the order of a licensed
practitioner
|
Electrical Stimulator
| Waveform |
dc "needle"
pulses with no parasitic spike
|
| Output Voltage |
0 to 500 peak V
|
| Output
Intensity |
20 µC maximum charge
per pulse into 100 Ohm
load
|
| Pulse Width |
narrow 10 µs, medium
20 µs, wide 30 µs
|
| Frequency Range |
1 to 160 pulses per s
|
| Surge On Duration,
Reciprocal On/Off Duration |
|
1, 2.5, 5, 7.5, 10, 15, 20, 30
s
| Surge Off
Duration |
2.5, 5, 10, 15,
20, 30, 50, 90 s
|
Instrument US/752
| Overall Dimensions |
19˝" W x 11"
D x 7" H
|
| Weight (including
transducer) |
20 lbs.
|
Ordering Information US/752
| Standard Soundhead |
32-US/752
|
| Standard QuickConnect
Soundhead |
32-US/752-D
|
| Small QuickConnect
Soundhead |
32-US/752-SM
|
| Standard/Small
QuickConnect Soundhead |
32-US/752-D2
|
| High Impact Plastic
Utility Cart |
4-APS-2000
|
| Flextrode® 8" x
10" Dispersive Electrode |
2-F106
|
|